Cardiac auscultation is one of the oldest diagnostic tools in medicine — and one of the hardest to digitize well. At PCARDIO Wearables, we develop MEMS microphone-based acoustic sensing systems for medical device OEMs building the next generation of digital stethoscopes, wearable cardiac monitors, and AI-driven auscultation platforms. This page walks through the full signal chain we engineer for our clients: from MEMS sensor selection and PDM-to-PCM conversion, through mechanical impedance matching, to segmentation-ready phonocardiogram (PCG) output. If you are scoping a heart sound analysis product, the engineering decisions below are the ones that will determine whether your device clears clinical validation.

Why MEMS Microphones Are Reshaping Digital Auscultation
Traditional electronic stethoscopes rely on piezoelectric transducers or electret condenser microphones. MEMS microphones offer OEMs three decisive advantages for cardiac acoustic sensing:
- Miniaturization: footprints below 4 × 3 mm enable integration into wearables, patches, and single-use disposables.
- Digital-native output: PDM interfaces eliminate analog front-end noise and simplify integration with low-power microcontrollers.
- Unit-to-unit consistency: MEMS process control delivers the reproducibility required for algorithm-based diagnostics and regulatory submission.
The trade-off is that MEMS microphones are designed for airborne acoustic pickup, not contact sensing against human tissue. Capturing clinically usable heart sounds — S1, S2, and the diagnostically critical murmurs in the 20–150 Hz band — requires deliberate engineering of the full acquisition chain.
The Phonocardiogram Signal — What We Are Actually Measuring
A phonocardiogram (PCG) captures the mechanical events of the cardiac cycle as pressure waves at the chest wall. Four signal features drive every downstream design decision:
- S1 — closure of mitral and tricuspid valves, marking the start of systole.
- S2 — closure of aortic and pulmonic valves, marking the start of diastole.
- Systolic and diastolic murmurs — turbulent flow indicating valvular or structural pathology (aortic stenosis, mitral regurgitation, VSD, aortic regurgitation, mitral stenosis, and others).
- Cycle timing — typical heart rates of 1–2 Hz define the envelope the segmentation algorithm must track.
Diagnostic value comes from the combination of timing (systolic vs. diastolic), shape (crescendo-decrescendo, plateau, decrescendo, decrescendo-crescendo), and auscultation location (aortic, pulmonic, tricuspid, mitral areas). A sensor system that cannot preserve all three dimensions is a sensor system that cannot support downstream classification — whether by a clinician or by a neural network.
Heart Sound Segmentation — Where Sensor Quality Meets Algorithm Performance
Segmentation is the process of parsing a PCG into S1, systole, S2, and diastole. It is the foundation on which all classification — rule-based or machine-learning-based — is built. Approaches in current literature include envelope detection, wavelet and Hilbert-Huang transforms, Hidden Markov Models, and logistic-regression-HSMM methods (Springer et al., 2016).
A recurring finding across the literature is that ECG-guided segmentation materially improves robustness: S1 follows the R-wave, S2 follows the T-wave, and co-acquisition allows multimodal neural networks to use the complementary timing information. For OEMs building a product that must work outside the quiet clinical environment — wearables, home monitoring, telemedicine — designing in a synchronized ECG channel is worth serious consideration at the architecture stage.
The segmentation algorithm is only as good as the envelope it receives. And the envelope is only as good as the sensor front end.
Classifying Heart Murmurs — The Three-Axis Framework OEMs Must Design Around
Any product claiming to detect or triage murmurs has to address three orthogonal classification axes:
Timing-based classification distinguishes systolic murmurs (aortic stenosis, mitral regurgitation, VSD, mitral valve prolapse) from diastolic murmurs (aortic regurgitation, pulmonic regurgitation, mitral stenosis) and continuous murmurs (patent ductus arteriosus).
Shape-based classification reads the temporal envelope: diamond-shaped (aortic and pulmonic stenosis), plateau or holosystolic (mitral and tricuspid regurgitation, VSD), decrescendo (aortic and pulmonic regurgitation), and decrescendo-crescendo (mid-diastolic mitral stenosis).
Location-based classification maps findings to anatomy: aortic area (2nd right ICS), pulmonic area (2nd left ICS), tricuspid area (4th left ICS), and mitral/apex (5th left ICS, midclavicular line), with Erb’s point (3rd left ICS) as a critical reference for aortic regurgitation and VSD.
No single axis is diagnostic on its own. A product that locks a clinician or algorithm into only one — for example, timing without location — will produce false positives. This has direct consequences for sensor count, sensor placement, and UI design in the end device.
Our Reference Acquisition Chain
PCARDIO’s reference design for heart sound acquisition is built around:
- Sensor: A MEMS microphone (digital I2C or PDM output), selected per our internal MEMS microphone selection guide for flat low-frequency response in the phonocardiographic band.
- Microcontroller: depending on whether the client prioritizes DSP headroom or wireless BLE connectivity.
- Acquisition parameters: 16 kHz sampling, 16-bit resolution, PDM filter gain — consistent across both MCU platforms for direct benchmarking.
- Transport: USB to host, with raw PCM streamed directly into MATLAB for real-time analysis during development.
The Mechanical Problem No Datasheet Warns You About
Early in this project we confirmed what every MEMS-based auscultation team eventually learns: direct contact between a MEMS microphone and skin does not produce a usable heart sound signal. The impedance mismatch between tissue and the microphone port attenuates the signal below the noise floor.
We iterated through three coupling strategies:
- Medical fabric over the port (fully bonded): blocked the acoustic aperture and attenuated signal to zero.
- Fabric bonded at edges only: thermally damaged the MEMS element during bonding, resulting in complete sensor failure. A reminder that MEMS microphones have strict reflow and thermal exposure limits that apply equally to post-assembly rework.
- Stethoscope tubing coupled to the microphone port: produced a clean, repeatable heart sound signal with clearly resolved S1 and S2 peaks.
This last result motivated our current mechanical-impedance-matching block — essentially a stethoscope bell acting as an acoustic transformer between the chest wall and the MEMS port. For OEMs, this raises three design questions we actively investigate for clients:
- Is the stethoscope membrane necessary? For digital acquisition, no — but it can improve SNR and reject ambient noise.
- Can the bell be replaced with a 3D-printed cone? Yes, with proper geometry and microphone positioning. This is the route we recommend for most wearable form factors.
- Does the cone need to be conical? The cone provides directional amplification and impedance matching, but a rectangular chamber can be acceptable if the signal processing chain is robust. The right answer depends on the product’s noise environment and target SNR.
Performance Targets and Verification Protocol
For any heart sound acquisition system we deliver, we verify against explicit, measurable criteria:
- Signal quality: S1 and S2 peaks clearly detectable; background noise below 25–50% of S2 peak amplitude.
- Morphological stability: S1/S2 pattern consistency across the full recording.
- Spectral content: majority of signal energy within the 20–150 Hz band.
- Frequency response: flat within 0.2 dB across 50, 70, 100, and 150 Hz, verified with a single-tone signal generator before any biological acquisition.
We evaluate candidate mechanical configurations — stethoscope with membrane, without membrane, 3D-printed cone, rectangular chamber — against these metrics on identical electronics, so design decisions are made on data rather than intuition.
What This Means for Your Product
If you are an OEM or startup scoping a heart sound analysis device, the engineering decisions that will determine your timeline and clinical performance are made early:
- MEMS microphone selection and its low-frequency response curve
- MCU architecture and whether you need co-acquired ECG
- The mechanical coupling block — this is the single biggest lever on signal quality
- The segmentation and classification algorithm architecture, and whether it must run on-device
PCARDIO / 2M Engineering has built the full stack described above and can take a concept from sensor selection through validated prototype to manufacturing-ready design. We work with medical device OEMs and startups across digital stethoscopes, wearable cardiac monitors, and AI-driven auscultation platforms